COMPANY HEADER
Request Information
Compact infrared camera for the short-wave range for non-contact temperature measurement of molten metal and metallic surfaces from 900 ºC to 2000 ºC
  • Highly dynamic CMOS detector with resolution up to 764 x 480 pixels
  • Special wavelength range from 500 nm to 540 nm minimizes errors due to unknown emissivity
  • Real-time analog output with 1 ms response time
Unit of Measure

Specifications

Optical Resolution at 32 Hertz (Hz) Frequency

N/A 764 x 480 px

Optical Resolution at 80 Hertz (Hz) Frequency (Switchable to 27 Hz)

N/A 382 x 288 px
Optical Resolution at 1 Kilohertz (kHz) Frequency1 N/A 72 x 56 px
Optical Resolution at 1 Kilohertz (kHz) Frequency (Fast Line-Scan Mode)2 N/A 764 x 8 px

Temperature Range at 27 Hertz (Hz) Modes

N/A 900 to 2000 ºC

Temperature Range at 80 and 32 Hertz (Hz) Modes

N/A 950 Up to 2000 ºC

Temperature Range at 1 Kilohertz (kHz) Modes

N/A 1100 Up to 2000 ºC

Spectral Range

N/A 500 to 540 nm
Frame Rate3 N/A Up to 1 kHz

System Accuracy (±)

N/A 1.5 % of Reading
Lens Field of View (FOV) at 764 x 480 Pixel (px) for f = 25 Millimeter (mm)4 N/A 26 x 16 º
Lens Field of View (FOV) at 382 x 288 Pixel (px) for f = 25 Millimeter (mm)5 N/A 13 x 10 º

Thermal Sensitivity (NETD) for 27, 32, and 80 Hertz (Hz) Frequency

N/A < 2 K 1400 ºC

Thermal Sensitivity (NETD) for 1 Kilohertz (kHz) Frequency

N/A < 2.5 K 1400 ºC

Detector

N/A CMOS

Detector Size

N/A 15 x 15 µm

Analog Output

N/A Optional Gigabit Ethernet (GigE) Universal Serial Bus (USB) 2.0

High-Speed Analog Output at 1 Kilohertz (kHz) Mode

N/A 1ms Real-Time Analog Output (0 – 10 V) of 8 x 8 Pixels (Freely Selectable Positions)

Input Voltage for Standard Process Interface (PIF)

N/A 0 to 10 V

Maximum Digital Input Voltage for Standard process interface (PIF)

N/A 24 V

Output Voltage for Standard Process Interface (PIF)

N/A 0 to 10 V

Industry Process Interface (PIF)

N/A 2x 0-10 V Inputs 3x 0(4)-20 mA Outputs 3x Relays (0-30 V/ 400 mA) Digital Input (Maximum 24 V) fail-safe relay
Universal Serial Bus (USB) Cable Length6 N/A 1 m (Standard) 10 m 5 m

Power Supply

N/A Universal Serial Bus (USB) Powered

Tripod Mount

N/A 1/4-20 UNC
Protection Class7 N/A Ingress Protection (IP67)

Ambient Temperature

N/A 5 to 50 ºC

Storage Temperature Range

N/A -40 to 70 ºC

Non-Condensing Relative Humidity

N/A 20 to 80 %

Vibration

N/A IEC 60068-2-27 (25 g and 50 g) IEC 60068-2-6 (sinus-shaped) IEC 60068-2-64 (broadband noise)

Housing Size

N/A 46 x 56 x 90 mm

Weight Including Lens

N/A 320 g

PC Requirements

N/A 1 GB RAM Minimum 1.5 GHz Windows XP SP 2 or Windows 7
Wide Measuring Range8 N/A 900 to 2000 ºC

Additional Information

Additional Information

N/A
  • Comprehensive software package and SDK
  • Ideally suitable for laser processing applications as radiation above 540 nm is blocked excellently

Software

Software

N/A
  • Display of the thermal image in real time with recording function (video, snapshot)
  • Complete set up of parameters and remote control of the camera
  • Detailed analysis of fast, thermodynamic processes
  • Output of analog temperature or alarm values via the process interface
  • Digital communication via RS232 or DLL for software integration
  • 1 Can be placed anywhere within the FOV
  • 2 Can be placed anywhere within the FOV
  • 3 1 ms real-time analog output (0 - 10 V) from 8 x 8 pixels (freely selectable)
    For fast processes
  • 4 Please note: measurement accuracy can be out of specification with distances below 500 mm
  • 5 Please note: measurement accuracy can be out of specification with distances below 500 mm
  • 6 5 m and 10 m also as high temperature USB cable (180 ºC)
  • 7 Only applies when lens protection tube is used.
  • 8 Without sub-ranges