Measuring Range |
N/A ±25 mm |
Linearity ≤±0.5 % Full Scale Output (FSO) |
N/A ≤±300 µm |
Linearity ≤±0.3 % Full Scale Output (FSO) |
N/A On Request |
Temperature Stability Full Scale Output (FSO) (Zero) |
N/A ≤70 ppm/K |
Temperature Stability Full Scale Output (FSO) (Maximum Temperature Error) |
N/A ≤150 ppm/K |
Sensitivity |
N/A 33 mV/mm/V |
Excitation Frequency |
N/A 1 kHz |
Excitation Voltage |
N/A 550 mV |
Connection |
N/A Axial cable outlet Cable diameter of 4.6 mm Integrated cable (3 m) with open ends Minimum bending radius 20 mm (fixed installation) |
Storage Temperature |
N/A -40 to 80 ºC |
Operating Temperature |
N/A -20 to 80 ºC |
Note for Operation Temperature |
N/A Optional up to 200 ºC on request |
Pressure Resistance |
N/A Atmospheric Pressure |
Note for Pressure Resistance |
N/A Optional 5 bar or 100 bar on front side on request. |
Shock (DIN-EN 60068-2-27) |
N/A 100 g40 g6 ms in 3 axes, 1000 shocks each 6 ms in 3 axes, 3 shocks each |
Vibration (DIN-EN 60068-2-6) |
N/A ±20 g±1.5 mm10 to 58 Hz in 2 axes, 10 cycles each 58 to 500 Hz in 2 axes, 10 cycles each |
Protection Class (DIN-EN 60529) |
N/A Ingress Protection (IP67) (Plugged) |
Housing Material |
N/A Stainless Steel |
Approximate Weight (Sensors) |
N/A 230 g |
Approximate Weight (Plunger) |
N/A 17 g |
Compatibility |
N/A MSC7401 MSC7602 MSC7802 |
Housing length L |
N/A 143.5 mm |
Plunger Length |
N/A 62 mm |
Houing Diameter |
N/A 20 mm |
Inner Diameter |
N/A 4.8 mm |
Plunger Diameter |
N/A 4 mm |
Characteristics |
N/A
|
Application |
N/A The DTA displacement sensors are suitable for numerous measurement tasks which require robust designs and high signal stability. Due to their wear-free design, the DTA sensors impress with longevity and long-term stability |
N/A
LVDT displacement sensors have a plunger which moves freely in the sensor housing. The plunger is joined to the object by a thread to transfer the movement of the measuring object. The measurement process in the sensor takes place without contact and is therefore wear-free. |