The new IMS5400-DS white light interferometer opens up new perspectives in industrial distance measurement. The controller has an intelligent evaluation feature and enables absolute measurements with nanometer accuracy at a relatively large offset distance. Compared to other absolute measuring optical systems, the IMS5400-DS offers an unsurpassed combination of accuracy, measuring range and offset distance.
Unit of Measure

Specifications

Measuring Range

N/A 2.1 mm

Approximate Start of Measuring Range

N/A 19 mm

Resolution

N/A <1 nm

Note for Resolution

N/A Measuring rate 0.5 kHz, moving averaging over 64 values, measured at the front of a glass plate in the mid of the measuring range (2 sigma).

Continuously Adjustable Measuring Rate

N/A 100 Hz to 6 kHz

Linearity

N/A <±50 nm

Note for Linearity

N/A Maximum deviation from reference system over the entire measuring range, measured on front surface of ND filter

Typical Temperature Stability Linearity

N/A 0.1 nm/K

Controller Temperature-Compensated Stability between 15 to 35 Degree Celsius (ºC) Temperature

N/A <10 ppm

Note for Typical Temperature Stability Linearity

N/A Without offset displacement

Light Source - Near Infrared Range (NIR) Superluminescent Light Emitting Diodes (SLED) Wavelength

N/A 840 nm

Laser Safety Class

N/A Class 1 in accordance with DIN EN 60825-1: 2015-07

Light Spot Diameter

N/A 10 µm

Note for Light Spot Diameter

N/A In the mid of the measuring range

Maximum Tilt Angle

N/A ±2 º

Note for Maximum Tilt Angle

N/A Maximum sensor tilt angle that produces a usable signal on polished glass (n = 1.5) in the mid of the measuring range. The accuracy decreases when approaching the limit values.

Target Material

N/A Diffuse Surfaces Glass Reflecting

Note for Target Material

N/A Non-transparent materials require an optically dense surface with a wavelength of 840 nm.

Direct Current (DC) Supply Voltage (± 15 Percent (%))

N/A 24 V

Approximate Power Consumption at 24 Volts (V) Voltage

N/A 10 W

Signal Input

N/A Sync in, Trigger in, 2 x encoders (A+, A-, B+, B-, index)

Digital Interface

N/A Ethernet/EtherCAT/RS422

Analog Output Current

N/A 4 to 20 mA

Analog Output Voltage

N/A 0 to 10 V

Digital to Analog (DA) Output

N/A 16 bits

Switching Output

N/A Error1-Out Error2-Out

Digital Output

N/A Sync out

Optical Connection

N/A Bending radius: static 30 mm, dynamic 40 mm pluggable optical fiber via E2000 socket (controller) and FC socket (sensor) Standard length 3 m, 5 m and 10 m; other cable lengths on request

Electrical Connection

N/A 11-pin I/O terminal strip (max. cable length 30 m) 3-pin output terminal strip (max. cable length 30 m) 3-pin supply terminal strip Encoder connection (15-pin, HD-sub socket, max. cable length 3 m, 30 m with external encoder supply) RJ45 socket for Ethernet (out)/EtherCAT (in/out) (max. cable length 100 m) RS422 connection socket (9-pin, Sub-D, max. cable length 30 m)

Sensor Installation

N/A Clamping Mounting Adapter

Controller Installation

N/A DIN Rail Mounting Free-Standing

Storage Temperature Range

N/A -20 to 70 ºC

Sensor Operation Temperature Range

N/A 5 to 70 ºC

Controller Operation Temperature Range

N/A 15 to 35 ºC

Shock (DIN-EN 60068-2-27)

N/A 6 ms in XY axis, 1000 shocks each 15 g

Vibration (DIN-EN 60068-2-6)

N/A 20 to 500 Hz in XY axis, 10 cycles each 2 g
Protection Class (DIN-EN 60529)1 N/A Ingress Protection (IP40)

Sensor Material

N/A Stainless steel

Controller Housing Material

N/A Aluminum, Passive Cooling

Control and Display Element

N/A 6 x color LEDs for intensity, range, SLED, pilot laser, status and power Multifunction button: two adjustable functions and reset to factory setting after 10 s Switchable pilot laser for sensor alignment (laser LED 635 nm, laser class 1, performance < 0.2 mW) Web interface for setup: selectable presets, freely selectable averaging possibilities, data reduction, setup management

Characteristics

Characteristics

N/A
  • Absolute measurement with nanometer accuracy, suitable for the measurement of e.g. step profiles
  • Compact and robust sensors with large offset distance
  • Measuring rate up to 6 kHz for high speed measurements
  • Ethernet/EtherCAT/RS422
  • Robust controller with passive cooling
  • Easy configuration via web interface

Additional Information

Additional Information

N/A Absolute distance measurement with nanometer resolution
The IMS5400-DS white light interferometer opens up new perspectives in industrial distance measurement. The controller has an intelligent evaluation feature and enables absolute measurements with nanometer accuracy at a relatively large offset distance. Compared to other absolute measuring optical systems, the IMS5400-DS offers an unsurpassed combination of accuracy, measuring range and offset distance.

Small light spot for the smallest of details and structures
The sensors generate a constantly small light spot over the entire measuring range. The light spot diameter is only 10 µm and allows the detection of small details such as structures on semiconductors and miniaturized electronic components.

Various interfaces for advanced automation
Integrated interfaces such as Ethernet, EtherCAT and RS422 as well as encoder connections, analog outputs, synchronization inputs and digital I/Os enable the connection to modern control systems and production programs.

Absolute measurement of step profiles
Unlike conventional interferometers, the IMS5400-DS also enables the measurement of step profiles. Thanks to the absolute measurement, the scanning is performed with high signal stability and precision. When measuring on moving objects, the differences in height of heels, steps and depressions can thus be reliably detected.

Integration in industrial environments
Robust sensors and a controller in a metal housing make the IMS5400-DS ideally suitable for integration into production lines. These compact sensors are extremely space saving and can also be integrated in confined spaces. The controller is installed in the control cabinet via DIN rail mounting and provides very stable measurement results due to active temperature compensation and passive cooling. Highly flexible fiber optic cables are available in lengths up to 20 m and allow a spatial separation of sensor and controller. Unlike other conventional interferometers, initial operations and parameter set up are easy and user-friendly via a web interface. No software installation is required.

Note

Note

N/A All data at constant ambient temperature (24 ±2 ºC)
  • 1 Controller and Sensor