Measurement Range |
N/A 25 mm |
Start of Measuring Range (SMR) |
N/A 25 mm |
Mid of Measuring Range (MMR) |
N/A 37.5 mm |
End of Measuring Range (EMR) |
N/A 50 mm |
Measuring Rate |
N/A 1.0 kHz10 kHz2 kHz4 kHz8 kHzContinuously Adjustable between 0.25 to 10 kHz250 Hz500 Hz |
Note for Measuring Rate |
N/A
7 Adjustable Stages Factory setting: measuring rate 4 kHz, median 9; modifying the factory setting requires the IF2001/USB converter. |
Linearity |
N/A <±5 µm |
Linearity Full Scale Output (FSO) |
N/A <±0.02 % |
Note for Linearity |
N/A Relates to digital output |
Repeatability |
N/A <0.8 µm |
Note for Repeatability |
N/A Typical value with measurements at 4 kHz and median 9 |
Temperature Stability Full Scale Output (FSO)1 | N/A ±0.005 %/K |
Light Spot Diameter - Start of Measuring Range (SMR) |
N/A 210 x 950 µm |
Light Spot Diameter - Mid of Measuring Range (MMR) |
N/A 80 x 970 µm |
Light Spot Diameter - End of Measuring Range (EMR) |
N/A 220 x 1000 µm |
Light Spot Diameter - Smallest Diameter |
N/A 80 x 970 µm with 37.5 mm |
Note for Light Spot Diameter |
N/A Light spot diameter with line-shaped laser determined based on the emulated 90/10 knife-edge method |
Light Source - Semiconductor Laser Wavelength |
N/A 670 nm<1 mW |
Light Source Color |
N/A Red |
Laser Safety Class |
N/A Class 2 in Accordance with DIN EN 60825-1: 2015-07 |
Permissible Ambient Light |
N/A 50,000 lux |
Direct Current (DC) Supply Voltage |
N/A 11 to 30 V |
Power Consumption at 24 Volts (V) Direct Current (DC) Voltage |
N/A <3 W |
Signal Input |
N/A 1 x HTL/TTL laser on/off 1 x HTL/TTL multi-function input: trigger in, slave in, zero setting, mastering, teach-in 1 x RS422 synchronization input: trigger in, sync in, master/slave, master/slave alternating |
Digital Interface |
N/A EtherNet IP (Connection via interface module) PROFINET (Connection via interface module) RS42 |
Digital Interface Data Rate |
N/A 18 bits |
Analog Output Current |
N/A 4 to 20 mA |
Analog Output Voltage |
N/A 0 to 5 V0 to 10 V |
Analog Output |
N/A 16 bit, freely scalable within the measuring range |
Switching Output |
N/A 2x switching outputs (error & limit value): npn, pnp, push pull |
Synchronization |
N/A Possible for simultaneous or alternating measurements |
Connection |
N/A Integrated cable 3 m, open ends, min. bending radius 30 mm (fixed installation) Integrated pigtail 0.3 m with 17-pin M12 plug Optional extension to 3 m / 6 m / 9 m / 15 m possible (suitable connection cable/optoNCDT catalog) |
Storage Temperature Range2 | N/A -20 to 70 ºC |
Operation Temperature Range3 | N/A 0 to 50 ºC |
Shock (DIN-EN 60068-2-29) |
N/A 15 g6 ms in 3 Axes |
Vibration (DIN-EN 60068-2-6) |
N/A 20 g20 to 500 Hz |
Protection Class (DIN-EN 60529) |
N/A Ingress Protection (IP67) |
Housing Material |
N/A Aluminum |
Approximate Weight with Pigtail |
N/A 185 g |
Approximate Weight with Cable |
N/A 300 g |
Control and Display Element |
N/A 2 x color LEDs for power/status Select & function keys: interface selections, mastering (zero), teach, presets, quality slider, frequency selection, factory settings web interface for setup (Connection to PC via IF2001/USB): application-specific presets, peak selection, video signal, freely selectable averaging possibilities, data reduction, setup management |
Characteristics |
N/A
|
N/A
The optoNCDT LL laser sensors produce an oval-shaped light spot of just a few millimeters in width which reminds of a short laser line (LL). This oval light spot compensates for inhomogeneous surface structures and the so-called speckling effect which particularly occurs with metallic surfaces. Combined with special software algorithms, interferences caused by surface roughness, defects, indentations or holes are filtered out. This enables stable and precise measurement results. |